PKKK/300/UP/016 HPLC Flexar Perkin Elmer
- Tapis semua fasa bergerak (akueus dan organik) menggunakan penuras membran 0.45 µm dan degas dalam kukus ultrasonik selama sekurang-kurangnya 15–20 minit.
- Lakukan Auto Purge / Manual Purge setiap kali pelarut ditambah atau ditukar jenis bagi menyingkirkan buih udara.
- Pastikan garisan dasar (baseline) dan tekanan pam stabil (RSD < 2%) sebelum memulakan suntikan sampel kelompok.
- Bagi sampel matriks kapsul lembut (softgel), gunakan Chloroform kerana gelatin/minyak tidak larut dalam Methanol.
|
NPRA PUSAT KAWALAN KUALITI |
ARAHAN KERJA: HPLC FLEXAR PERKIN ELMER | |
| No. Dokumen: PKKK/300/UP/016 | Terbitan / Semakan: Terbitan 3 Semakan 0 | |
| Tarikh Kuatkuasa: 10 April 2026 | Bahagian / Unit: SPPK · Unit Penyaringan | |
0.0 SEJARAH PINDAAN & SEMAKAN
Sejarah Pindaan & Semakan Dokumen (1 entri) ▸
Terbitan 2, Semakan 0:
Sejarah Pindaan & Semakan Dokumen (1 entri) ▸
Terbitan 2, Semakan 1:
Sejarah Pindaan & Semakan Dokumen (3 entri) ▸
Terbitan 2, Semakan 2:
Mengemaskini perenggan 5.3, penolong pegawai farmasi ditukar daripada U36 / U32 kepada U7 / U6.
Terbitan 3, Semakan 0:
Kemas kini tarikh kuat kuasa dari 1 Julai 2025 kepada 10 Apr 2026.
Pinda Laporan Penyelenggaraan, Servis dan Kalibrasi di dalam fail alat berkenaan kepada Rekod penyelenggaraan alat Shimadzu High Performance Liquid Chromatography Flexar (Perkin Elmer)
1.0 TUJUAN
2.0 SKOP
6.0 PROSEDUR
Pengenalan alat
This instrument consists of the following components:
FLEXAR LC Pumps
FLEXAR PDA Plus LC Detector
FLEXAR LC Column Ovens
FLEXAR Autosamplers
Computer
Printer
Operasi alat
Switch ON the battery
Double click <TC Navigator> to enter Total Chrom
At this moment, “Hands-on” button grey-out
Starting the Total Chrom Navigator
Seizing – establishing software & HPLC communication
Seizing – Communicating (L) and Communication established. Hands on available (R)
Preparation
To Start Pump – Click instrument setup
Browse & Choose a sequence to trigger pump
Browse sequence
A sequence displayed 🡪 Action 🡪 Setup
Status: Setting up
Status: HPLC Ready / Pump Off 🡪 Click Hands On
Hands-On 🡪 Pump (or Purge)
Pump – Adjust the composition 100% of mobile phase.
Key in flow rate 3-5mL/min
Repeat 1 – 6 for other line if necessary
Tighten purge valve and remove
Syringe after completed.
Connect the column to the inlet and outlet tubing (could be not connected)
The flow rate and % composition of mobile phases could be change
Equilibrate the column from by slowly increasing flow rate to 1mL/min or as required by method
Benchmark of column C18 (150mm x 4.6mm x 5µm)
At 1mL of 75% MeOH 25% UPW, pressure redback is 2250psi
Hands On 🡪 Autosampler 🡪 Flush Autosampler (perform a few times to get rid of air bubbles in syringe)
Double click Oven.exe
Key-in temperature 🡪 Enter
End of preparation: Run 🡪 Clear Setup
End of preparation: Run 🡪 Clear Setup 🡪 Status
Status: No Method – ready for use
Sequence
Running sample – Click “Sequence”
Browse or select sequence / Create New
Specify the Sample Type, Vial Number, Instrument Method
Double click the Data column to browse the directory (folder to save raw file) and name the raw file.
Example of directory and raw files (xxx.raw)
Highlight rows to run, or just click Actions 🡪 Set Up
HPLC Status: Setting Up – Method parameters being send to HPLC (wait a while)
Status of pre-injection & injection
Ready – Equilibrium (green) (L) and Pump start / equilibrating flow rates (R)
Status – Flushing & Running
Flushing loop, syringe, injecting & running (blue)
Details – view current run sample
Method
Overview of Method Editor (Only run instrument)
Overview of Method Editor (with calibration & processing Method)
Autosampler – Set Injection volume & temperature
Pump configuration
Pump Program (Isocratic)
Pump Program (Gradient)
Summary of Gradient Mode
Instrument Timed Events – Leave blank 🡪 OK. Save the Method Editor
Processing Method Integration & Method
External Calibration
Chromatogram of Data001 will be displayed. Notice that this data file is linked to this method (eg. Test1)
To optimize the processing method, click <Process><Noise/Area Threshold>
You will notice that most of the noises will not be integrated
Go to <Edit Components…> to name the components.
Move the box to the first peak of your interest and type the name of the compound in <Name> field. Click Next. Continue naming the peaks in similar manner.
To open the method file, click <Other><Method>
The method file is opened. Now you can see your component name and retention time appeared in the component list.
Go to <Component><Edit Component>
Click OK
Go to <Component><Calibrate…>
Highlight the data file and select the appropriate level. Click Add.
Click OK
You can see other’s calibration by highlighting the next component.
Save the method by clicking the save icon
Totalchrom – Format Report Editor
A report format is opened with same name as method file name (.rpt)
Go to Report and select whichever column you like to include as well as the column number
Example here showed how to insert column 2 with Component Name
You must insert Raw Amount column if you have done your calibration. This column is indicating the amount / concentration directly calculated from the calibration curve.
Go to Option
To insert the chromatography plot between header and report body, tick <Print replot with reports>
Build a sequence to reprocess stored data
To build a sequence using a template
Method Editor 🡪 Other 🡪 Sequence Editor
Sequence Editor 🡪 File 🡪 New
Sequence Editor 🡪 File 🡪 New 🡪 Global Parameters
A sequence displayed 🡪 Double click “Data” to input .raw files for each row 🡪 Browse & load the correct raw files for each calibration level & samples.
Sequence Editor 🡪 Action 🡪 Batch
Ensure starting & ending row no. set correctly 🡪 Set the rest as following
Batch reprocessing 🡪 Automatically proceed or click the button as showed in the picture
Browse the result file one by one (.rst file)
File 🡪 Print Preview Report
Print Preview Report
System Suitability
TotalChrom Navigator 🡪 Apps 🡪 Suitability
Option 🡪 Suitability Option
Signal to Noise (Peak Height vs. baseline)
Choose the SST criteria 🡪 Input the limits 🡪 OK
File 🡪 Save As 🡪 OK 🡪 Enter File Name (*.sui)
Results Files Selection 🡪 Choose a Result File (.rst file)
The report displayed the criteria that user has chosen
Print report as usual
View / Print Chromatogram
View chromatograms 🡪 Click Overlay icon
Overlay Chromatograms 🡪 2D
Options 🡪 3-D Preview
Adjust X and Y axis 🡪 OK
3-D Chromatogram 🡪 File 🡪 Print 🡪 Check criteria to print
Summary: Repeatability of multiple injections
Double click on the summary button
File List 🡪 Select one or more result files (.rst) 🡪 Add
Files to summarize – displayed the selected .rsst files 🡪 OK
Load from Result File (or Method)
File 🡪 Open Format 🡪 Select a Summary template (.sum file)
Summary interface – grey turned to white
IRIS – Processing software
Browse the first .rst file for viewing spectra
Main View : Chromatogram, contour, spectra & 3d view
Use Compare Tab
Browse and open the 2nd .rst file for comparison
Compare 2 Chromatogram side by side
Overlay chromatogram
Right click on one of the chromatogram, select ”Overlay Chromatogram”
Stack View of Chromatogram & Overlay Spectras
Wavelength Maximum
Move the cursor on spectra to read and label the optimum wavelengths.
Print chromatogram & overlay spectras
File 🡪 Print 🡪 Choose printer (or PDF) 🡪 Choose All Panels in view
File 🡪 Save Results
File 🡪 Print 🡪 All Pane 🡪 OK
3.0 DEFINISI
HPLC – High Performance Liquid Chromatography
4.0 CARTA ALIRAN
Tiada carta aliran untuk prosedur ini.
5.0 TANGGUNGJAWAB
Ketua Unit / Pegawai Farmasi Bertanggungjawab
Memastikan prosedur ini diikuti semasa pengendalian alat ini.
Pegawai Farmasi
Mengendalikan alat ini mengikut prosedur yang telah ditetapkan.
Menyelia proses pengendalian alat ini oleh Penolong Pegawai Farmasi.
Penolong Pegawai Farmasi U7 / U6
Memastikan penyelenggaraan alat ini dijalankan pada waktu yang ditetapkan dan menyimpan rekod penyelenggaraan alat.
Penolong Pegawai Farmasi
Mengendalikan alat ini mengikut prosedur yang telah ditetapkan.
7. REKOD KUALITI
Rekod penyelenggaraan alat Shimadzu High Performance Liquid Chromatography
Flexar (Perkin Elmer)
Jadual Kromatografi & Rujukan Data
| Terbitan | Semakan | Ditulis Oleh | Disemak Oleh | Diluluskan Oleh | Tarikh Kuatkuasa |
|---|---|---|---|---|---|
| 1 | 0 | Lim Wen Jun | Nur Aziah Mohd Rasali | Ooi Suat Hong | 1 Jan 2021 |
| 2 | 0 | Nur Aziah Mohd Rasali | Ooi Suat Hong | Wan Nurul Aina Mior Abdullah | 1 Nov 2022 |
| 2 | 1 | Joanne Ong Yen Nee | Ooi Suat Hong | Nur Aziah Mohd Rasali | 1 April 2024 |
| 2 | 2 | Tan Lu Yi | Ooi Suat Hong | Ida Syazrina Ibrahim | 1 Julai 2025 |
| 3 | 0 | Nurul Nadiah Noh | Tan Lu Yi | Ida Syazrina Ibrahim | 10 Apr 2026 |
| RUJUKAN | RUJUKAN |
|---|---|
| No. Dokumen | Tajuk |
| Manual Operasi Alat |
| Flow rate (mL/min) | Mobile Phases | Pressure (psi) |
|---|---|---|
| 1 | 75% MeOH 25% UPW | 2250 |
| 2 | 75% MeOH 25% UPW | 4320 |