PKKK/300/UP/022 Identifikasi Anti diabetik dalam produk tradisional menggunakan HPLC
- Tapis semua fasa bergerak (akueus dan organik) menggunakan penuras membran 0.45 ยตm dan degas dalam kukus ultrasonik selama sekurang-kurangnya 15โ20 minit.
- Lakukan Auto Purge / Manual Purge setiap kali pelarut ditambah atau ditukar jenis bagi menyingkirkan buih udara.
- Pastikan garisan dasar (baseline) dan tekanan pam stabil (RSD < 2%) sebelum memulakan suntikan sampel kelompok.
- Bagi sampel matriks kapsul lembut (softgel), gunakan Chloroform kerana gelatin/minyak tidak larut dalam Methanol.
|
NPRA PUSAT KAWALAN KUALITI |
ARAHAN KERJA: IDENTIFIKASI ANTI DIABETIK DALAM PRODUK TRADISIONAL MENGGUNAKAN HPLC | |
| No. Dokumen: PKKK/300/UP/022 | Terbitan / Semakan: Terbitan 3 Semakan 0 | |
| Tarikh Kuatkuasa: 10 April 2026 | Bahagian / Unit: SPPK ยท Unit Penyaringan | |
0.0 SEJARAH PINDAAN & SEMAKAN
Sejarah Pindaan & Semakan Dokumen (13 entri) โธ
Terbitan 1, Semakan 1:
Kemaskini 6.6 System suitability test requirement, di mana
kritieria Theoretical plate telah dikemaskini daripada โ(NLT) 1500 kepada โ(NLT) 2000โ.
Menambah โ%RSD for retention time: Not more than, NMT 2.0%โ
Menambah perenggan 6.7 Internal Quality Control
Mengemaskini perenggan 5.4, penolong pegawai farmasi ditukar daripada U36 / U32 kepada U7 / U6
Mengemaskini perenggan 3 Definisi
Menambah perenggan 6.3.4 Single Standard Stock Solution
Mengemaskini perenggan 6.4.2 dengan menukar Weigh kepada Measure untuk sampel cecair
Menambah perenggan 6.8 Acceptance criteria for test sample
Mengemaskini perenggan 6.5.5 Injection Sequence
Mengemaskini perenggan 6.9.7 Evidence of analysis report
Terbitan 2, Semakan 0:
PKKK/300/UAT/004 kepada PKKK/300/UP/003
PKKK/300/UAT/012 kepada PKKK/300/UP/011
PKKK/300/UAT/017 kepada PKKK/300/UP/016
PKKK/300/UAT/001 kepada PKKK/300/UP/001
PKKK/300/UAT/009 kepada PKKK/300/UP/008
PKKK/300/UAT/010 kepada PKKK/300/UP/009
Sejarah Pindaan & Semakan Dokumen (3 entri) โธ
Mengemaskini para 7 Rekod Kualiti dengan menambah:
UP/001A Borang Persampelan
UP/005 Laporan Pengujian HPLC
1.0 TUJUAN
Untuk memastikan ujian identifikasi Anti- diabetik menggunakan alat High Performance Liquid Chromatography (HPLC) dilaksanakan dengan cekap dan berkesan.
2.0 SKOP
6.0 PROSEDUR
Materials and Equipments
High Performance Liquid Chromatography with photodiode array (PDA) detector
pH Meter
Metformin
Glipizide
Gliclazide
Glibenclamide
Ammonium Dihydrogen Orthophosphate
Sodium Heptane Sulfonate
Deionized water
Phosphoric Acid
Acetonitrile
Beakers (50 ml)
Solvent filter system
Nylon solvent filter 0.45 ยตm
Solvent bottle (2L and 1L)
Amber Volumetric Flasks (20 ml)
Spatula
Disposable syringe without needle (5 ml)
2 ml glass screw cap vial
Preparation of Mobile Phase
Measure 1 L of acetonitrile.
Degassed filtered buffer mobile phase and acetonitrile for at least for 20 minutes to ensure no bubble present.
Preparation of Reference Standard
Individual standard stock solution
Concentration of each standard stock: 0.5 mg/ml (500 ppm)
Pipette 0.8 ml of each standard stock solution into 20 ml amber volumetric flask. Dilute to volume with methanol and mix well.
Concentration of each standard in mixed solution: 0.02 mg/ml (20 ppm)
Pipette 2.5 ml of mix working standard solution into 5 ml amber volumetric flask. Dilute to volume with Methanol (0.01 mg/ml or 10 ppm).
6.3.4.1 Prepare a single standard stock solution following the same procedure outlined in 6.3.1 to 6.3.3, but using a single standard instead of multiple standards.
6.3.4.2 This procedure is applicable for screening selected compounds only.
Preparation of sample
Sample handling (sampling)
Sampling process differ between different type of dosage form. Please refer to document PKKK/300/UP/001 (Proses Persampelan).
Sample preparation for analysis
For solid sample: Weigh 1 gm of the sample in a beaker and dissolve with 20 ml methanol.
For liquid sample: Measure 10ml of the sample in a beaker and dissolve with 10 ml methanol.
Experimental Method
The HPLC system used for identification of anti-diabetic are given in Table 1 below.
Table 1: HPLC system used for identification of anti-diabetic
Mobile Phase Method
Solvent gradient program for mobile phase is use in identification of anti-diabetic shown as below:
Table 2: Gradient program for mobile phase
Refer to document PKKK/300/UP/003 (RRLC 1200 Series), PKKK/300/UP/011 (HPLC Prominence-i) and PKKK/300/UP/016 (HPLC Flexar) on how to set-up method and HPLC system.
Standard Injection
Sample Injection
Retention time and UV spectrum observation
To observe and check retention time and UV spectrum of the sample and RS, refer to document PKKK/300/UP/003 (RRLC 1200 Series), PKKK/300/UP/011 (HPLC Prominence-i) and PKKK/300/UP/016 (HPLC Flexar).
First, inject a blank or diluent, then inject mix working standard solution six times (SST).
If more than five samples are injected in a single analysis, the LOD solution must be injected once every five samples.
A blank or diluent must be injected before the next sample injection. Refer to Table 3 for an example of the sequence table:
Table 3: Injection sequence for identification of anti-diabetics in HPLC
LOD solution must be injected at the end of the analysis.
System suitability test requirement
System suitability test need to be performed for each batch run to ensure the entire system is working properly. Perform at least six replicate injections of mix working standard solution (0.02 mg/ml).
Acceptance criteria:
%RSD for retention time: Not more than, NMT 2.0%
Tailing factor not more than (NMT) 2.0
Theoretical plate not less than (NLT) 2000
Internal Quality Control (IQC)
The percentage difference of retention time for LOD solution should not be more than 3 % for every batch of 5 samples.
Acceptance criteria for test sample
Analysis report
Each sample will have one analysis report. Report must consist of detail as below:
Sample details (Sample name, sample number and batch number)
Analysis date
Equipment parameter including equipment name, mobile phase program and HPLC system.
List and preparation procedures of RS use in analysis
Sample preparation procedure and printed sample weight (for solid sample only).
Interpretation of result (Detected, Not Detected or suspected of specific anti-diabetic compound)
Evidence of analysis must be attached together in analysis report:
Chromatogram of 1st and 2nd injection of sample (with minimum information of compound name, peak area and retention time)
If sample sequence is more than 5 samples, a standard chromatogram will be injected again after 5 sample injections set and presented along with report with the nearest standard chromatogram injection.
Overlay spectrum of sample that have similar retention time with RS.
IQC table
3.0 DEFINISI
4.0 CARTA ALIRAN
Tiada carta aliran untuk prosedur ini.
5.0 TANGGUNGJAWAB
Ketua Seksyen
Memastikan prosedur ini dipatuhi semasa pengujian dijalankan.
Menyemak dan mengesahkan semua laporan ujian tidak lulus yang telah disemak oleh Pegawai Farmasi termasuk laporan OOS.
Ketua Unit / Pegawai Farmasi Bertanggungjawab
Memastikan prosedur ini dipatuhi semasa pengujian dijalankan.
Menyemak dan mengesahkan semua laporan ujian yang telah disemak oleh Pegawai Farmasi termasuk laporan OOS.
Mengagihkan sampel kepada Pegawai Farmasi dan Penolong Pegawai Farmasi.
Menyelia ujian yang dijalankan oleh Pegawai Farmasi dan Penolong Pegawai Farmasi.
Menyemak laporan ujian.
Menjalankan siasatan OOS dan menyediakan laporan OOS bagi sampel gagal ujian.
Menjalankan aktiviti pengujian sampel mengikut tatacara yang telah ditetapkan.
Merekodkan bilangan ujian selepas selesai laporan ujian disiapkan untuk tujuan laporan bulanan.
Menghantar laporan ujian ke Unit Perkhidmatan Analisis untuk simpanan.
Menjalankan aktiviti pengujian sampel mengikut tatacara yang telah ditetapkan
7. REKOD KUALITI
No. Borang Tajuk
UP/001A Borang Persampelan
UP/005 Laporan Pengujian HPLC
8.0 LAMPIRAN
UV Spectrum of Anti-Diabetic Standard
Metformin
Glipizide
Gliclazide
Glibenclamide
Jadual Kromatografi & Rujukan Data
| Terbitan | Semakan | Ditulis Oleh | Disemak Oleh | Diluluskan Oleh | Tarikh Kuatkuasa |
|---|---|---|---|---|---|
| 1 | 0 | Joanne Ong Yen Nee | Ooi Suat Hong | Ida Syazrina Ibrahim | 15 Nov 2024 |
| 1 | 1 | Tan Lu Yi | Ooi Suat Hong | Ida Syazrina Ibrahim | 1 Julai 2025 |
| 2 | 0 | Nurul โIzzah Hany Ismail | Tan Lu Yi | Ida Syazrina Ibrahim | 10 April 2026 |
| RUJUKAN | RUJUKAN |
|---|---|
| No. Dokumen | Tajuk |
| PKKK/300/UP/003 | Rapid Resolution Liquid Chromatography HP 1200 |
| PKKK/300/UP/011 | Shimadzu High Performance Liquid Chromatography (Prominence-i) |
| PKKK/300/UP/016 | High Performance Liquid Chromatography Flexar (Perkin Elmer) |
| PKKK/300/UP/001 | Proses Persampelan (Ujian Penyaringan) |
| PKKK/300/UP/008 | pH Meter Model FiveEasy Plus |
| PKKK/300/UP/009 | Alat Timbang Metler Toledo ME204T |
| Column | Zorbax Eclipse Plus โ C18 or equivalent |
|---|---|
| Dimensions | 250 x 4.6 mm x 5 ยตm |
| Flow rate | 1.2 mL/min |
| Column temperature | 30oC |
| Autosampler Temperature | 4oC |
| Wavelength detector (PDA) | 234 nm |
| Injection volume | 10 ยตL |
| Run time | 16 minutes |
| Mobile Phase Method | Gradient program |
| Time (min) | Buffer 0.01M Ammonium Dihydrogen Phosphate Buffer + 0.06% Sodium Heptane Sulfonate pH 3.3 (%) | Acetonitrile (%) |
|---|---|---|
| 0 | 95 | 5 |
| 1 | 95 | 5 |
| 2 | 85 | 15 |
| 10 | 25 | 75 |
| 13 | 25 | 75 |
| 13.1 | 95 | 5 |
| 16 | 95 | 5 |
| No. | Detail | No. of Injection |
|---|---|---|
| 1 | Blank / Diluent | 1 |
| 2 | Mix working standard solution (SST) | 6 |
| 3 | LOD solution | 1 |
| 4 | Blank / Diluent | 1 |
| 5 | Sample number 1 | 2 |
| 6 | Blank / Diluent | 1 |
| 7 | Sample number 2 | 2 |
| 8 | Blank / Diluent | 1 |
| 9 | Sample number 3 | 2 |
| 10 | Blank / Diluent | 1 |
| 11 | Sample number 4 | 2 |
| 12 | Blank / Diluent | 1 |
| 13 | Sample number 5 | 2 |
| 14 | LOD solution | 1 |
| 15 | Blank / Diluent | 1 |
| 16 | Sample number 6 | 2 |
| 17 | Blank / Diluent | 1 |
| 18 | LOD solution | 1 |