PKKK/300/UP/026 Identifikasi Anti-hipertensi dalam produk tradisional menggunakan HPLC

Arahan Kerja Rasmi (Level 300 SOP) Ā· Kategori: HPLC / LC
1
1. Reagen & Piawai
Penyediaan fasa bergerak, stok & kalibrasi
āž”
2
2. Pengekstrakan
Timbang, sonikasi, sentrifug & turas 0.45µm
āž”
3
3. Persediaan Alat
Purging saluran, autotune & baseline
āž”
4
4. Suntikan Batch
Blank, SST (n=6), Sampel & IQC
āž”
5
5. Verifikasi & Rekod
Padanan RT, spektrum UV/MS & Laporan
⚔ Spesifikasi & Parameter Kromatografi / Analitikal (Official Parameters):
Instrumen UtamaHPLC Shimadzu Prominence-i / Agilent 1200
Turus / ColumnZorbax Eclipse XDB-C18 (150 Ɨ 4.6 mm, 5 µm)
Fasa Bergerak / PelarutBuffer Fosfat pH 3.0 : Methanol : Acetonitrile
Kadar Alir & Suhu1.0 mL/min @ 40 °C
Pengesan / Gelombang / SIM238 nm (Amlodipine, Atenolol, Captopril, Losartan, Nifedipine)
Kriteria Kesesuaian Sistem (SST)SST %RSD ≤ 2.0%, Tailing T ≤ 2.0, resolusi puncak Rs > 1.5.
šŸŽÆ Sebatian Sasaran Pengujian:
DexamethasoneBetamethasonePrednisonePrednisoloneTriamcinolone acetonideHydrocortisone acetateCortisone acetateBetamethasone-17-valerate
šŸ’” Petua & Perhatian Penting Makmal (Bench Notes):
Analyst Benchtop Checklist Progress
0 / 28 Langkah Selesai (0%)
NPRA
PUSAT KAWALAN KUALITI
ARAHAN KERJA: IDENTIFIKASI ANTI-HIPERTENSI DALAM PRODUK TRADISIONAL MENGGUNAKAN HPLC
No. Dokumen: PKKK/300/UP/026 Terbitan / Semakan: Terbitan 3 Semakan 0
Tarikh Kuatkuasa: 10 April 2026 Bahagian / Unit: SPPK Ā· Unit Penyaringan
šŸ“‹

0.0 SEJARAH PINDAAN & SEMAKAN

šŸ“‹ Sejarah Pindaan & Semakan Dokumen (14 entri) ā–ø

Terbitan 1, Semakan 1:

Kemaskini 6.6 System suitability test requirement, di mana

kritieria Theoretical plate telah dikemaskini daripada ā€œ(NLT) 1500 kepada ā€œ(NLT) 2000ā€.

Menambah ā€œ%RSD for retention time: Not more than, NMT 2.0%ā€

Menambah perenggan 6.7 Internal Quality Control

Mengemaskini perenggan 5.4, penolong pegawai farmasi ditukar daripada U36 / U32 kepada U7 / U6

Mengemaskini perenggan 3 Definisi

Menambah perenggan 6.3.4 Single Standard Stock Solution

Mengemaskini perenggan 6.4.2 dengan menukar ā€œWeighā€ kepada ā€œMeasureā€ untuk sampel cecair

Menambah perenggan 6.8 Acceptance criteria for test sample

Mengemaskini perenggan 6.5.5 Injection Sequence

Mengemaskini perenggan 6.6.7 dengan menambah ā€œIQC tableā€

Terbitan 2, Semakan 0:

Pindakan nombor dokumen PKKK/300/UAT/052 kepada PKKK/300/UP/026.

PKKK/300/UAT/004 kepada PKKK/300/UP/003

PKKK/300/UAT/012 kepada PKKK/300/UP/011

PKKK/300/UAT/017 kepada PKKK/300/UP/016

PKKK/300/UAT/001 kepada PKKK/300/UP/001

šŸ“‹ Sejarah Pindaan & Semakan Dokumen (3 entri) ā–ø

Mengemaskini para 7 Rekod Kualiti dengan menambah:

UP/001A Borang Persampelan

UP/005 Laporan Pengujian HPLC

šŸŽÆ

1.0 TUJUAN

Untuk memastikan ujian identifikasi anti-hypertensives menggunakan alat High Performance Liquid Chromatography (HPLC) dilaksanakan dengan cekap dan berkesan

šŸ“

2.0 SKOP

šŸ”¬

6.0 PROSEDUR

Materials and Equipments

Instrument:

High Performance Liquid Chromatography with photodiode array (PDA) detector

Analytical & microbalance

Column: Zorbax SB C18 5µm, 250 mm x 4.6 mm or equivalent

Reference standards: Primary or secondary RS or equivalent

Captopril

Methyldopa

Atenolol

Metoprolol

Labetalol

Nifedipine

Prazosin

Chemicals:

Methanol

Acetonitrile

Deionized water

Sodium dihydrogen orthophosphate

Glasswares and consumables:

Beakers (50 ml)

Solvent filter system

Nylon solvent filter 0.45 µm

Solvent bottle (1 L)

Amber Volumetric Flasks (Grade A)

Nylon syringe filter 0.45 µm

Disposable syringe without needle (5 ml)

2 ml amber glass screw cap vial

Preparation of Mobile Phase

Buffer preparation:

To prepare sodium dihydrogen orthophosphate dihydrate (25mM) buffer pH 3.0, dissolve 3.9 g of sodium dihydrogen orthophosphate dihydrate in 1L of deionized water.

Measure 1 L of acetonitrile and transfer in another solvent bottle.

Degassed filtered mobile phase at least for 20 minutes to ensure no bubble present.

Preparation of Reference Standard

Individual standard stock solution

Combine mix working standard solution by transfer 0.8 ml of captopril stock solution and 0.4 ml of each of the other reference standard solutions into one amber volumetric flask (10 ml) and make up to volume with methanol. Mix well.

6.3.4.1 Prepare a single standard stock solution following the same procedure outlined in 6.3.1 to 6.3.3, but using a single standard instead of multiple standards.

6.3.4.2 This procedure is applicable for screening selected compounds only

Preparation of sample

Sample handling (sampling)

Sampling process differ between different type of dosage form. Please refer to document PKKK/300/UP/001 (Proses Persampelan).

Sample preparation for analysis

For solid sample: Weigh 0.5 gm of the sample in a beaker and dissolve with 20 ml methanol.

For liquid sample: Measure 10 ml of the sample in a beaker and dissolve with 10 ml methanol.

ā–¹ 2) Sonicate solution for 30 minutes.
ā–¹ 3) Centrifuge at 4500 rpm for 15 mins
ā–¹ 4) Filter supernatant layer with 0.45 µm syringe filter into 2 mL vial.

Experimental Method

The HPLC system used for identification of anti-hypertensives are given in Table 1 below:

Table 1: HPLC system used for identification of anti-hypertensive

Mobile Phase Method

Solvent gradient program for mobile phase is use in identification of anti-hypertensives shown as below:

Table 2: Gradient program for mobile phase

Refer to document PKKK/300/UP/003 (RRLC 1200 Series), PKKK/300/UP/011(HPLC Prominence-i) and PKKK/300/UP/016 (HPLC Flexar) on how to set-up method and HPLC system.

Standard Injection

Sample Injection

Retention time and UV spectrum observation

To observe and check retention time and UV spectrum of the sample and RS, refer to document PKKK/300/UP/003 (RRLC 1200 Series), PKKK/300/UP/011 (HPLC Prominence-i) and PKKK/300/UP/016 (HPLC Flexar).

Injection Sequence

First, inject a blank or diluent, then inject mix working standard solution six times (SST).

If more than five samples are injected in a single analysis, the LOD solution must be injected once every five samples.

A blank or diluent must be injected before the next sample injection. Refer to Table 3 for an example of the sequence table:

Table 3: Injection sequence for identification of anti-hypertensive in HPLC

v. LOD solution must be injected at the end of the analysis.

System Suitability Test (SST) requirement

System suitability test needs to be performed for each batch run to ensure the entire system is working properly. Perform at least six replicate injections of mix working standard solution.

Acceptance criteria:

%RSD for retention time: Not more than, NMT 2.0%

Tailing factor not more than (NMT) 2.0

Theoretical plate not less than (NLT) 2000

Internal Quality Control (IQC)

The percentage difference of retention time for LOD solution should not be more than 3 % for every batch of 5 samples.

Acceptance criteria for test sample

ā–¹ a) Retention time of the compound in the sample chromatogram corresponds to the retention time of the same compound in the standard chromatogram.
ā–¹ b) Peak area of the compound is above than peak area of the LOD concentration.
ā–¹ c) The UV spectrum is identical to the spectrum of the standard with similarity of more than 0.999 to the spectrum of standard.

Analysis report

Each sample will have one analysis report. Report must consist of detail as below:

Sample details (Sample name, sample number and batch number)

Analysis date

Equipment parameter including equipment name, mobile phase program and HPLC system.

List and preparation procedures of RS use in analysis.

Sample preparation procedure and printed sample weight (for solid sample only).

Interpretation of result (Detected, Not Detected or suspected of specific anti-hypertensives compound).

Evidence of analysis must be attached together in analysis report:

Chromatogram of 1st and 2nd injection of sample (with minimum information of compound name, peak area and retention time)

Chromatogram of 1st and 2nd injection of standard (with minimum information of compound name, peak area and retention time). If sample is sequence at 6 or more, chromatogram of standard before sample must be presented

Overlay spectrum of sample that have similar retention time with RS

IQC table

šŸ“–

3.0 DEFINISI

šŸ”€

4.0 CARTA ALIRAN

Tiada carta aliran untuk prosedur ini.

šŸ‘„

5.0 TANGGUNGJAWAB

Ketua Seksyen

Memastikan prosedur ini dipatuhi semasa pengujian dijalankan

Menyemak dan mengesahkan semua laporan ujian tidak lulus yang telah disemak oleh Pegawai Farmasi termasuk laporan OOS

Ketua Unit / Pegawai Farmasi Bertanggungjawab

Memastikan prosedur ini dipatuhi semasa pengujian dijalankan

Menyemak dan mengesahkan semua laporan ujian yang telah disemak oleh Pegawai Farmasi termasuk laporan OOS

Mengagihkan sampel kepada Pegawai Farmasi dan Penolong Pegawai Farmasi

Menyelia ujian yang dijalankan oleh Pegawai Farmasi dan Penolong Pegawai Farmasi

Menyemak laporan ujian

Menjalankan siasatan OOS dan menyediakan laporan OOS bagi sampel gagal ujian

Menjalankan aktiviti pengujian sampel mengikut tatacara yang telah ditetapkan

Merekodkan bilangan ujian selepas selesai laporan ujian disiapkan untuk tujuan laporan bulanan

Menghantar laporan ujian ke Unit Perkhidmatan Analisis untuk simpanan

Menjalankan aktiviti pengujian sampel mengikut tatacara yang telah ditetapkan

šŸ“

7. REKOD KUALITI

No. Borang Tajuk

UP/001A Borang Persampelan

UP/005 Laporan Pengujian HPLC

šŸ“Ž

8.0 LAMPIRAN

šŸ“Ž

8.0 LAMPIRAN

UV Spectrum of Anti-hypertensives Standard

šŸ“Ž

8.0 LAMPIRAN

šŸ“Ž

8.0 LAMPIRAN

šŸ“Ž

8.0 LAMPIRAN

šŸ“Š

Jadual Kromatografi & Rujukan Data

SEJARAH SEMAKAN PKKK/300/UP/026
TerbitanSemakanDitulis OlehDisemak OlehDiluluskan OlehTarikh Kuatkuasa
10Nurul ā€˜Izzah Hany IsmailOoi Suat HongIda Syazrina Ibrahim15 Nov 2024
11Tan Lu YiOoi Suat HongIda Syazrina Ibrahim1 Julai 2025
20Nurul ā€˜Izzah Hany IsmailTan Lu YiIda Syazrina Ibrahim10 April 2026
JADUAL 2 PKKK/300/UP/026
RUJUKANRUJUKAN
No. DokumenTajuk
PKKK/300/UP/003Rapid Resolution Liquid Chromatography HP 1200
PKKK/300/UP/011Shimadzu High Performance Liquid Chromatography (Prominence-i)
PKKK/300/UP/016High Performance Liquid Chromatography Flexar (Perkin Elmer)
PKKK/300/UP/001Proses Persampelan (Ujian Penyaringan)
JADUAL 3 PKKK/300/UP/026
ColumnZorbax SB C18 or equivalent
Dimensions250 x 4.6 mm x 5 µm
Flow rate1.0 ml/min
Column temperature40oC
Autosampler temperature4oC
Wavelength detector (PDA)208 nm: Captopril 225 nm: Methyldopa, Atenolol, Metoprolol, Labetalol 235 nm: Nifedipine 250 nm: Prazosin
Injection volume10 µl
Run time30 minutes
Mobile Phase MethodGradient program
JADUAL 4 PKKK/300/UP/026
Time (min)Buffer (%)Acetonitrile (%)
0.01 – 16.99955
17.005050
17.01 - 25.001585
25.01 - 30.00955
JADUAL 5 PKKK/300/UP/026
No.DetailNo. of injection
1Blank / Diluent1
2Mix working standard solution (SST)6
3LOD solution1
4Blank / Diluent1
5Sample number 12
6Blank / Diluent1
7Sample number 22
8Blank / Diluent1
9Sample number 32
10Blank / Diluent1
11Sample number 42
12Blank / Diluent1
13Sample number 52
14LOD solution1
15Blank / Diluent1
16Sample number 62
17Blank / Diluent1
18LOD solution1